Is My Instrument Accurate Enough for This?#

Matching instrument specs to measurement needs. A 3½-digit DMM is perfect for checking a 5V power rail and useless for measuring a 1.000V reference to four digits. Knowing when the instrument is the limiting factor — and when it’s more than adequate — prevents both false confidence and unnecessary upgrades.

DMM Resolution and Accuracy#

Resolution is the smallest increment the display can show. Accuracy is how close the reading is to the true value. They’re different things.

DMM typeDisplay countsResolution (on 4V range)
3½-digit20001 mV
4½-digit20,0000.1 mV
5½-digit200,0000.01 mV
6½-digit2,000,0000.001 mV

Accuracy specification: Typically stated as ±(% of reading + counts)

Example: ±(0.5% + 3 counts) measuring 5.000V = ±(0.025V + 0.003V) = ±0.028V

So the true voltage is somewhere between 4.972V and 5.028V.

MeasurementRequired accuracy3½-digit DMM sufficient?
Is 5V rail present? (±10%)±0.5VYes
Is 3.3V rail within spec? (±5%)±0.165VYes
Matching 1% resistors±0.5% or betterMarginal
Checking 2.500V reference to ±0.1%±2.5 mVNo — need 4½ digits

Oscilloscope Accuracy#

Most bench oscilloscopes have 3–5% vertical accuracy — vastly worse than a DMM. The scope shows waveshape; the DMM tells the actual voltage.

MeasurementUse scope?Use DMM?
DC voltage to ±1%NoYes
Waveform shape, edges, ringingYesNo
Peak-to-peak on a fast signalYes (only option)N/A

Sample rate: Nyquist rule requires at least 2× the highest frequency component. For accurate waveform reconstruction, 5–10× is needed.

Signal typeMinimum sample rateRecommended
1 MHz square wave (~3 ns edges)600 MS/s2 GS/s
1 ns rise time2 GS/s5–10 GS/s

When the Instrument Isn’t Enough#

Signs of hitting instrument limits:

  • Reading fluctuates in last digit more than expected from signal
  • Trying to see a difference smaller than accuracy spec
  • Scope displays clean square wave but ringing is suspected below bandwidth
  • FFT noise floor higher than signal being sought

Options:

  • DMM accuracy insufficient → use 4½/5½-digit DMM or check against known reference
  • Scope bandwidth too low → higher-bandwidth probe/scope or infer from slower measurements
  • Noise floor too high → average, use bandwidth limiting, improve probe grounding

Tips#

  • DMM accuracy degrades at extremes of each range — measuring 0.5V on a 40V range wastes most resolution
  • Use the lowest range that accommodates expected voltage
  • Scope timing accuracy (~50 ppm) is much better than amplitude accuracy (~3–5%)

Caveats#

  • Accuracy specs assume 23°C ±5°C and within calibration period — outside these, accuracy degrades
  • AC accuracy is much worse than DC accuracy on most DMMs, and degrades with frequency
  • Sample rate per channel may be lower than headline rate when using multiple channels
  • Memory depth limits effective sample rate at long timebases
  • Averaging reduces random noise but hides intermittent events

In Practice#

  • Measurement that fluctuates in last two digits when signal should be stable indicates instrument resolution/accuracy limit
  • Two instruments disagreeing on same measurement indicates at least one is at its accuracy limit — trust the one with better specs
  • Reading that seems wrong but is within accuracy spec may actually be correct — the spec defines expected uncertainty
  • Attempt to measure smaller difference than accuracy spec is meaningless — get a better instrument